Skip to main content
Fig. 5 | Journal of Therapeutic Ultrasound

Fig. 5

From: Rapid full-wave phase aberration correction method for transcranial high-intensity focused ultrasound therapies

Fig. 5

Model pressure patterns. The left column displays the pressure patterns at the geometric focus through the aberrator model obtained from the hydrophone scans. The right column displays the results of the simulated pressure patterns using the acoustic parameters listed in Table 1. The rows from top to bottom represent the cases of uncorrected, corrected using simulated phases calculated with known acoustic parameters, and corrected using phases obtained from hydrophone measurements, respectively

Back to article page