Skip to main content

Table 1 Imaging protocols

From: Volumetric MR-guided high-intensity focused ultrasound ablation to treat uterine fibroids through the abdominal scars using scar patch: a case report

Examination

Sequence

Aim

Imaging plane

TR (ms)

TE (ms)

FA (°)

Section thickness (mm)

Field of view (mm)

Matrix size

TA (s)

Screening

3D T2 TSE

Uterine fibroid

Sagittal

1500

165

90

3

250 × 250

176 × 173

177

 

FFE

Scar visualization

Coronal

3.4

1.74

10

2

200 × 200

172 × 173

79

 

CE-T1 TSE*

Uterine fibroid

Coronal

5.5

2.7

12

2.5

250 × 250

192 × 192

130

Treatment

3D T2 TSE

Treatment planning

Sagittal

1550

150

90

1.6

250 × 250

176 × 173

124.9

 

FFE

Scar and scar patch

Coronal

3.4

1.74

10

2

200 × 200

172 × 173

79

 

EPI

Temperature mapping

Coronal, sagittal

37

19.5

19

7

400 × 400

160 × 160

2.9

 

CE-T1 TSE*

Non-perfused volume

Coronal

5.5

2.7

12

2.5

250 × 250

192 × 192

130

6-month follow-up

3D T2 TSE

Uterine fibroid

Sagittal

1487

165

90

2.5

250 × 250

176 × 173

140

 

CE-T1 TSE*

Non-perfused volume

Coronal

5.5

2.7

12

2.5

250 × 250

192 × 192

130