Optical measurement of skin temperature in MR-HIFU

Background/introduction

MR-guided high-intensity focused ultrasound (MR-HIFU) treatments may cause skin heating in the vicinity of the treatment site. Current MR thermometry methods do not provide reliable measurements of skin temperature either during the sonication or during the cool-down periods between sonications. These technical challenges require additional pauses to decrease the likelihood of skin burns, thus impacting treatment duration. Therefore, quantitative, accurate, and rapid techniques are needed to measure surface skin temperature during HIFU treatment. This study aims to develop an optical method that detects temperature changes at the skin surface to maintain a safe skin temperature during treatment and to reduce pauses between sonications.

Methods

Chiral nematic liquid crystal slurry (LCR Hallcrest, LLC, Glenview, IL, #09-NSL33), which displays a change in reflected light color between 25-30°C was mixed with 0.4% w/v agarose gel at 80-85°C to create a 4 mL mixture of 10% liquid crystal slurry.

The mixture was poured onto a black-painted glass microscope slide and then covered with another glass slide (gel thickness = 0.5mm). Nickel-plated steel wire (220μm-thick) was used to heat the slide. Electrical current through the wire was adjusted to ensure a steady-state gradient across the glass slide that spanned from below 25°C to above 30°C. Color change of the liquid crystal gel layer was imaged using a DSLR camera and a stable light source placed at a 57° angle. A 0.2-mm diameter thermocouple (Omega Engineering, Inc., Stamford, CT, USA) sampled temperature at different distances from the heating wire to measure the temperature gradient. Local pixel color intensities around the thermocouple tip were isolated and spatially matched with measured temperatures using a custom MATLAB routine (Mathworks, Inc., Naticks, MA, USA). Multivariate polynomial regression was performed on the data to determine temperature as a function of red, green, and blue intensities (3rd order polynomials used for each).

Results and conclusions

Color of the reflected light from the thermochromic liquid crystals varied with temperature in the agarose gel. As the temperature increased within the gel, red, green, and blue color intensities increase and level off at different rates. A multivariate polynomial regression showed RMS error of 0.74°C for the fit. Our data show that thermochromic liquid crystals provide an accurate, dynamic, and absolute method to monitor skin temperature during MR-HIFU treatment and ensure patient safety. While the liquid crystals used herein displayed a change of color in the 25-30°C range, crystals with other color-temperature ranges are commercially available and will be evaluated in further studies.

figure 1

Figure 1

Acknowledgements (Funding)

Funding was provided by the Sheikh Zayed Institute for Pediatric Surgical Innovation at Children’s National Medical Center and the Princeton Internships in Civic Service program.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yang, D., Celik, H., Wackerle, D. et al. Optical measurement of skin temperature in MR-HIFU. J Ther Ultrasound 3 (Suppl 1), P81 (2015). https://doi.org/10.1186/2050-5736-3-S1-P81

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2050-5736-3-S1-P81

Keywords