Skip to main content

Table 1 Validation of the proposed motion compensation framework under clinical conditions. Percentage of the volume due for ablation estimated to have received a lethal amount of thermal dose with (first row) and without (second row) the proposed motion compensation framework enabled

From: A framework for continuous target tracking during MR-guided high intensity focused ultrasound thermal ablations in the abdomen

 

Animal #1

Animal #2

Animal #3

With motion correction

88%

74%

70%

Without motion correction

36%

63%

40%