Table 2 (abstract O34). Implemented Sequence Performance Parameters. tSNR efficiency is proportional to (δxyz*√(Tseq)*σT)−1 . For Monitoring - Background, median slice uncertainty reported

From: International Society for Therapeutic Ultrasound Conference 2016

  Speed (s) FOV (cm) Slices Res (mm) Slice Thickness Uncertainty (phantom) Uncertainty (in vivo) in vivo tSNR Efficiency
2DFT 3.82 28 1 1.09×2.19 3 MM 0.48°C 0.66°C 100%
Focal Spot Localization 7.00 36 5 1.10×1.10 3 mm 0.16°C 0.38°C 253%
Monitoring - Focus 2.39 36 3 2.00×2.00 2 mm 0.15°C 0.41°C 182%
Monitoring - Boundary 4.78 36 2 2.00×2.00 2 mm 1.03°C 0.33°C 160%
Monitoring - Background 9.56 36 24 2.00×2.00 2 mm 0.15°C 0.54°C 69%