Skip to main content

Table 1 Values and uncertainties of k

From: Uncertainty estimation for temperature measurement with diagnostic ultrasound

  Final k Standard deviation Gaussian uncertainty Uncertainty
Discrete frame-to-frame 748.0 °C 76.6 °C (10.2 %) 35.4 °C (4.8 %) 112.0 °C (15.0 %)
Continuous frame-to-base 747.9 °C 80.1 °C (10.7 %) 12.3 °C (1.7 %) 92.4 °C (12.4 %)
  1. Standard deviation was calculated between all k in the measurement, Gaussian uncertainties of the values of k were averaged, and (final) uncertainty was obtained by adding both uncertainties