Fig. 7

From: Open-source, small-animal magnetic resonance-guided focused ultrasound system

Fig. 7

Sonications at 1.1 (a) and 3.68 (b) MHz FUS frequencies, targeting temperature set points in ROI 1 for 10 min. Background phase drifts were corrected using an ROI outside of the area of heating (ROI 2). Controller voltage is also plotted for each case and also stabilizes after an initial rise and small overshoot. The white arrow indicates surface coil placement. Low-temperature SNR at the top of the phantoms (and far from the surface coil which sat at the level of the water-phantom interface) contributed to the apparent elevated temperatures there but did not interfere with the focus measurements. Stripe artifacts in the water cone are likely due to Moire fringes caused by poor field homogeneity in the water bath near the transducer causing aliasing