Fig. 10 | Journal of Therapeutic Ultrasound

Fig. 10

From: Open-source, small-animal magnetic resonance-guided focused ultrasound system

Fig. 10

ARFI-measured displacement in the axial and coronal directions overlaid on T 1-weighted images of a tofu phantom. Localized displacements are apparent at the focus while smoother displacements appear throughout which may correspond to shear waves. The focal displacement location and size correspond well to the expected geometry of the transducer. Acoustic reflections with the air boundary and lower SNR near the top of the phantom likely contribute to the diffuse rise in displacement there. Moire fringes are visible at the top of the coronal image, which are likely due to field inhomogeneity near the surface of the transducer causing some water bath signals to be excited and alias into the image