From: Deep Bleeder Acoustic Coagulation (DBAC)—part II: in vivo testing of a research prototype system
Exp. no. | Side | Bleed type | Bleed depth (mm) | Depth category | Flow rate (ml/min) | Flow rate category | Bleed character correct? |
---|---|---|---|---|---|---|---|
1 | L | SFA | 49 | Shallow | 104 | Fast | Y |
2 | L | DFA | 115 | Deep | N/A | ||
3 | R | SFA | 52 | Shallow | 95 | Fast | Y |
4 | R | DFA | 78 | Deep | N/A | ||
5 | R | SFA | 52 | Shallow | 24 | Slow | Y |
6 | R | SFA | 52 | Shallow | 24 | Slow | Na |
7 | R | DFA | 105 | Deep | N/A | ||
8 | L | SFA | 92 | Deep | 12.5 | Slow | Y |
9 | L | SFA | 92 | Deep | 83 | Fast | Y |
10 | L | DFA | 110 | Deep | N/A | ||
11 | R | DFA | 110 | Deep | N/A | ||
12 | R | SFA | 56 | Shallow | 5.2 | Slow | Y |
13 | R | SFA | 56 | Shallow | 5.2 | Slow | Na |
14 | L | DFA | 108 | Deep | N/A | ||
15 | L | SFA | 52 | Shallow | 88 | Fast | Y |