Fig. 5 | Journal of Therapeutic Ultrasound

Fig. 5

From: Characterization and evaluation of tissue-mimicking gelatin phantoms for use with MRgFUS

Fig. 5

Thermal repeatability for the 250-bloom phantom, demonstrating the degree of consistency of achieving the same measured peak temperature for a given acoustic power. For clarity in this figure, only some representative runs are included. The order of applied power was 6.6, 6.6, 6.6, 10.1, 6.6, 13.2, 6.6, 13.2, 6.6, 15.4, 6.6, 17.6, 6.6, 20.7, 20.7, and 6.6 W (italicized values reflect the runs shown in the figure and listed in the legend)