Table 1 Sonication parameters used in the facet joint phantom experiment

From: Magnetic resonance-guided focused ultrasound treatment of facet joint pain: summary of preclinical phase

Sonication number

Energy (J)

Temperature at the focal point (°C)

Average temperature in the vicinity of the intervertebral foramen (°C)

1

300

61

38

2

300

65

38

3

300

73

38

4

300

84

37

5

300

62

38

6

300

59

39

7

300

66

37

8

300

65

38

9

300

71

39

10

600

83

38

11

600

67

37

12

900

72

39

13

900

75

42

14

1,200

78

43

15

1,200

64

41

  1. Summary of the sonication parameters used in the phantom experiment, including acoustic energy delivered, the temperature measured at the focal point on the targeted facet joint, and the temperature measured in the nontargeted intervertebral foramina, located in the far field of the acoustic beam.