Figure 10From: Prediction and suppression of HIFU-induced vessel rupture using passive cavitation detection in an ex vivo modelComparison of controller-inactive and controller-active trials in rupture suppression experiments. Means and standard errors are shown for (A) time to rupture from start of sonication, (B) trial-average SPTP acoustic intensity at the focus, and (C) total acoustic energy delivered.Back to article page