Table 3 Clot fragmentation (post-/pre-wet filter weight) 11-μm filter

From: Transcranial sonothrombolysis using high-intensity focused ultrasound: impact of increasing output power on clot fragmentation

Group Acoustic output power (W) Number Mean weight (mg) Standard deviation Min (mg) Median (mg) Max (mg) P Value
1 0 60 2.4 0 −8.2 1.8 11.8 -
2 50 62 3.3 0 −8.2 2.8 19.8 >0.9999
3 100 20 18.5 0.07 −2.2 3.3 310.8 >0.9999
4 125 63 2.7 0 −6.2 2.8 13.8 >0.9999
5 150 65 4.2 0.01 −6.2 2.8 50.8 >0.9999
6 200 20 3 0 −3.2 2.8 12.8 >0.9999
7 235 20 2.2 0 −2.2 1.8 5.8 >0.9999
8 270 22 2.2 0 −3.2 2.3 9.8 >0.9999
9 400 20 2.7 0 −1.2 1.8 9.8 >0.9999
Overall - 352 3.9 0.02 −8.2 2.8 310.8 -
  1. p Values reflect comparison to the 0 W group, adjusted for multiple comparison.