Skip to main content

Table 1 Total acoustic powers and I SA calculated at corresponding input electric power levels

From: Estimating dynamic changes of tissue attenuation coefficient during high-intensity focused ultrasound treatment

Input electric power (W) Total acoustic power (W) ISA(W/cm2)
11 5 117
70 34 737
75 37 801
80 39 845
90 44 961
100 49 1,068