Destruction of atherosclerotic plaque using pulse ultrasound with a planar rectangular transducer
Journal of Therapeutic Ultrasound volume 3, Article number: O72 (2015)
Background/introduction
The aim of this paper is to present a feasibility study for using mechanical mode ultrasound for destroying atherosclerotic plaque with MRI monitoring.
Methods
An MRI safe flat rectangular (3x10 mm2) transducer operating at 5 MHz was used. A spatial average temporal average intensity of 10 W/cm2 for 120 s was used, with DF of 10 % and 1 ms pulse repetition period. This optimized protocol was applied in a rabbit model. The plaque in the rabbit was created using a 2 % cholesterol diet.
Results and conclusions
During the course of the diet, the aorta of the rabbit was imaged using high resolution T2 W FSE every one month in order to assess the progress of the growth of the plaque. Selected rabbits were sacrificed in one, two, and three months since the initiation of the diet in order to assess the growth of the plaque histologically. Approximately 50 % of the artery was covered by plaque in 3 months since the initiation of the diet. This was confirmed using MRI and histology. The proposed mechanical protocol has successfully destroyed plaque in the aorta of the rabbit. The size of the rabbit aorta has similar size with the coronary or carotid arteries in humans. The rabbit is a good model for creating atherosclerotic plaque.
Acknowledgements (Funding)
This work was supported by the Research Promotion Foundation (RPF) of Cyprus (program ΕΠΙΧΕΙΡΗΣΕΙΣ/ΠΡΟΪΟΝ/0311/01, and the European regional development structural funds).
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Damianou, C., Christofi, C., Mylonas, N. et al. Destruction of atherosclerotic plaque using pulse ultrasound with a planar rectangular transducer. J Ther Ultrasound 3 (Suppl 1), O72 (2015). https://doi.org/10.1186/2050-5736-3-S1-O72
Published:
DOI: https://doi.org/10.1186/2050-5736-3-S1-O72